Features *Design for medical & lab instruments *Application for centrifuges *Universal input: 90~264Vac *Over voltage protection *Over current protection *Short circuit protection *Meet EN60950, UL1950, and CSA 22.2 safety standard *Meet FCC part 15 class B, EN55022, class B, EN61000-<2,3,4 EN60555-2 EMC *requirements *100% burn-in *Compact size 192x100x65mm |
|
Specifications General |
|
Operating: 0~40℃ |
|
Temperature |
|
Storage: -20~+85℃ |
|
Temperature |
|
Input Input Voltage: 90~264Vac Universal input Input current: 3A/115Vac, 1.5A/230Vac Input Frequency: 47~63 Hz |
|
Output |
|
Output Rating +5V / 0.5A: 0A min. 0.8A peak +32V/3.4A: 0A min. 3.9A peak |
|
Line Regulation: 0.5% typical,1% maximum Load Regulation: main output 2.0% <<<<<<<<<<<<.<<Aux. output 5.0% |
|
Output Voltage Tolerance: +32V main output << <<<<< +4V adjustable |
|
Ripple & Noise: Maximum 1% peak-peak Over Voltage<<<Protection: 130% 10% on the output |
|
Ripple & Noise: Maximum 1% peak-peak Over Voltage Protection: 130% 10% on the output |
|
Over Load Protection: The over load protection will shut down PSU when total power exceeds 150% of maximum power at input voltage 120Vac |
|
Short Circuit Protection: The short circuit protection will shut down PSU on any output shorted and automatic recovery without any damage while fault condition is removed |
|
Mechanical |
|
Dimension: Length 192.0mm Width: 100.0mm Height: 65.0mm Weight: 0.83Kg |